Click here for the V6 Multi-Channel Syringe Pump
Click here for the Syringe V6 Multi-Channel 55 Series Product Line

V6 Multi-Channel
Syringe Pump


The V6 55 Series Multi-Channel Pump is a programmable precision fluid control device. It utilizes a non-volatile user program memory and system input/output (I/O). The V6 Multi-Channel Pump is designed with the flexibility and versatility to meet a variety of system configurations. It can be used as a stand alone controller, or integrated into system configurations. The module is capable of simple closed loop control of an external quantity or process without the need of an external controller for direction.

Mechanically, the syringe pump is driven via stepper motor and precision ball screw. It has a 60mm stroke and is available in two resolutions. Pump communications are compatible with other Kloehn devices.

This V6 55 Series Multi-Channel Pump uses Non-Standard length syringes.
For additional custom modifications, see your Kloehn Sales Representative.


Physical
• Overall Height 11.83”
• Maximum Width with an 8 Channel Manifold 6.5” (minimum width with a 2 Channel Manifold 3.2 “)
• Overall Depth 5.75”
• Weight 6.0 lbs.

Environmental
• Operating temperature of 0 to 50 C
• Operating humidity of 5 to 95% RH, non-condensing at 50 C
• Storage temperature of –25C to 85 C
• RoHS and WEEE compliant

Mechanical
• Mounting holes at top and bottom as well as on the face plate
• Resolutions available (24000 and 48000 increments)
• Maximum Speed is 10000 steps / second.
• Syringe sizes: 25ul to 5ml (supplied separately)
• Valve type: 3 Way Solenoid Valve each channel
• PC board: Driver & Control boards (fully functional motor drivers and control electronics)
• Card-edge connector
• Integrated RS232 Integrated power filter

Interface
• RS-485 or RS-232 communications
• Data terminal or OEM protocol
• 1200 to 38.4K baud, 8 data bits, 1 stop bit, no parity, half duplex
• Three inputs, three output ports
• Digital voltmeter
• Switched and wired device address

Power
• Operation at 24V (with a maximum power voltage ripple of 720mv peak-to-peak) Power consumption – 80 watts, 11 watts idle.

Syringe Pump:
V6 55 Series Multi-Channel Pump


48K
RESOLUTION
24K
RESOLUTION
NUMBER OF
CHANNELS
PUMP PART
NUMBER
DECAL
(OPTIONAL)
X 8 CHANNEL 29514 29555
X 7 CHANNEL 29515 29555
X 6 CHANNEL 29516 29555
X 5 CHANNEL 29517 29555
X 4 CHANNEL 29518 29555
X 3 CHANNEL 29519 29555
X 2 CHANNEL 29520 29555
X 8 CHANNEL 29524 29555
X 7 CHANNEL 29413 29555
X 6 CHANNEL 29411 29555
X 5 CHANNEL 29409 29555
X 4 CHANNEL 29399 29555
X 3 CHANNEL 29397 29555
X 2 CHANNEL 29395 29555

(Add the decal to the order as required)

Ancillary items:

DESCRIPTION PART NUMBER
Communication cable RS485, 6” long; pump to pump (2 units) 17736 or 17736X
Cooling Fan Option 24290
Starter Kit (includes all items listed below) 24682
• 24VDC Power supply 24680
• Adapter (card edge to AMP connectors) 23428
• Software; operator’s manual; and application notes. 23422
• Communications cable (RS232 / 5’ long; pump to computer) 17734
Connector (card edge with solder pins) 26875

Software support:
• KloehnControl (configuration and debug program)
• KSerial (command-line communication utility)
• KCom (machine language helper program)
• KProbe (communication trouble shooting utility)

Valve—3 Way Mini-Solenoid:
• Part number 80841
• Orifice Diameter: .040”
• Internal Volume: 42ul
• Zero Dead Volume: No un-swept paths
• Rated Pressure: Vacuum to 40psi
• Wetted Materials: Peek, FFKM

Standard Teflon Syringe Assembly:
• Thread Type: 1/4-28 UNF-2A
• Glass Type: Borosilicate glass
• Individually Boxed
• Fine Tuning Adjustment: Zero Dead Volume Only
• * Tip nose is 316L Stainless Steel Teflon coated

Size Orifice Standard Syringe Zero Dead Volume
(ZDV)
25ul .016 29896 29898*
50ul .024 29435 29856*
100ul .032 29437 29858
250ul .039 29439 29441
500ul .076 29442 29443
1.0ml .076 29447 29448
1.25ml .076 29451 29450
2.5ml .076 29454 29452
5.0ml .076 29456 29458

UHMW-PE Syringe Assembly:
• Thread Type: 1/4-28 UNF-2A
• Glass Type: Borosilicate glass
• Individually Boxed
• Fine Tuning Adjustment: Zero Dead Volume Only
• * Tip nose is 316L Stainless Steel Teflon coated

Size Orifice Standard Syringe Zero Dead Volume
(ZDV)
25ul .016 29897 29899*
50ul .024 29436 29857*
100ul .032 29438 29859
250ul .039 29440 29861
500ul .076 29444 29863
1.0ml .076 29445 29446
1.25ml .076 29864 29449
2.5ml .076 29455 29453
5.0ml .076 29457 29459




Syringe V6 Multi-Channel 55 Series Product Line

The Syringe V6 Multi-Channel 55 Series line is designed to couple to a v6 Multi-Channel 55 Series Pump and standard valve assembly. A fine tuning design feature is incorporated in the Zero Dead Volume (ZDV) syringes for the 25ul to the5.0ml. These syringes are typically used in various types of instruments and analyzers serving clinical, medical, and analytical instrumentation applications. An array of syringe volumes are available that can be dispensed and/or aspirated over a full 6cm stroke.

For additional custom modifications, see your Kloehn Sales Representative.

Physical
• Length (Dispensed): 4.610”
• Thread Type: 1/4-28 UNF-2A
• Test Pressure: 80 psi

Environmental
• Operating Temperature of 10 to 40°C
• Storage Temperature of -25 to 85°C
• Relative Humidity: Up to 100%

Mechanical

Syringe Sizes 25ul 50ul 100ul 250ul 500ul 1.0ul 1.25ul 2.5ul 5.0ul
Orifice Diameter (in) .027 .024 .032 .039 .076 .076 .076 .076 .076
Syringe Diameter (in) .04 .04 .04 .04 .04 .04 .04 .05 .06
*Max Drag Force (lbs) PTFE 3.0 3.0 3.0 3.0 3.0 3.0 3.0 3.0 6.0
UHMW 7.0 7.0 7.0 7.0 7.0 7.0 7.0 7.0 10.0

* Add 1 pound drag force to non-lubricated plunger tip requests.

Chemical
• Standard Syringe Seal: PTFE
• Wetted Materials: Borosilicate Glass, Kel-F, PTFE (or UHMW) (smaller sizes ZDV tips are Teflon coated 316L)
• Syringe tips lubricated with laboratory-grade Silicone