The VersaPump 6 is a programmable precision fluid control device. It utilizes a non-volatile user program memory and system input/output (I/O). The V6 is designed with the flexibility and versatility to meet a variety of system configurations. It can be used as a stand alone controller, or integrated into system configurations. The module is capable of simple closed loop control of an external quantity or process without the need of an external controller for direction.
Mechanically, the syringe pump is driven via stepper motor and precision ball screw. It has a 60mm stroke and is available in three resolutions. The pump can be coupled with a variety of valves and syringes, as well as PC board selections, to meet specific needs. Pump communications are compatible with other Kloehn devices.
Physical
• Height 10”
• Width 2.55”
• Depth 4.75” (face plate to back plate) 4.9” to card edge
• Weight 5.14 lbs
Environmental
• Operating temperature of 0 to 50 C
• Operating humidity of 5 to 95% RH, non-condensing at 50 C
• Storage temperature of 25C to 85 C
• RoHS & WEEE compliant
Mechanical
• Operation in any orientation
• Mounting holes at top and bottom as well as on the face plate
• Resolutions available (12000; 24000; and 48000 increments)
• Speed; (Max. - 10000 steps / sec.; min. 60 steps / sec.) (Default - 5000 steps / sec.)
• Syringe size - 25ul to 50ml
• Valve type 2 way to 12 way
• 3 PC board choices:
Base board (sensor array and motor pin out only - no motor drivers or control electronics).
Driver board (sensor array and motor drivers only, no control electronics).
Driver & Control boards (fully functional motor drivers and control electronics).
• Card-edge connector
• Integrated RS232 Integrated power filter
• Multiple valve choices, including no valve & solenoid valve
|
1/10 Stroke Dispenses |
Full Stroke Dispenses |
| Precision: |
0.10% CV, 0.20% max. |
0.007% CV, 0.0092% max. |
| Accuracy: |
0.20% CV, 0.60% max. |
-------- |
Interface
• RS-485 or RS-232 communications
• Data terminal or OEM protocol
• 1200 to 38.4K baud, 8 data bits, 1 stop bit, no parity, half duplex
• Three inputs, three output ports
• Digital voltmeter
• Switched and wired device address
Power
• Operation at 24V (with a maximum power voltage ripple of 720mv peak-to-peak) Power consumption 40 watts, 11 watts idle.
Syringe Pump:
|
| RESOLUTION |
VALVE FEATURE |
DRIVER & CONTROL FEATURE |
PUMP PT# |
DECAL PT# |
| 48K |
24K |
12K |
NONE |
SOLE-
NOID |
ROTARY |
BASIC |
DRIVER
BOARD |
DRIVER &
CONTROL BOARD |
| X |
|
|
X |
|
|
X |
|
|
54000 |
23279 |
| X |
|
|
X |
|
|
|
X |
|
54001 |
23279 |
| X |
|
|
X |
|
|
|
|
X |
55002 |
23279 |
| X |
|
|
|
X |
|
X |
|
|
54010 |
23279 |
| X |
|
|
|
X |
|
|
X |
|
54011 |
23279 |
| X |
|
|
|
X |
|
|
|
X |
55012 |
23279 |
| X |
|
|
|
|
X |
X |
|
|
54020 |
23279 |
| X |
|
|
|
|
X |
|
X |
|
54021 |
23279 |
| X |
|
|
|
|
X |
|
|
X |
55022 |
23279 |
|
X |
|
X |
|
|
X |
|
|
54100 |
23279 |
|
X |
|
X |
|
|
|
X |
|
54101 |
23279 |
|
X |
|
X |
|
|
|
|
X |
55102 |
23279 |
|
X |
|
|
X |
|
X |
|
|
54110 |
23279 |
|
X |
|
|
X |
|
|
X |
|
54111 |
23279 |
|
X |
|
|
X |
|
|
|
X |
55112 |
23279 |
|
X |
|
|
|
X |
X |
|
|
54120 |
23279 |
|
X |
|
|
|
X |
|
X |
|
54121 |
23279 |
|
X |
|
|
|
X |
|
|
X |
55122 |
23279 |
|
|
X |
X |
|
|
X |
|
|
54200 |
23279 |
|
|
X |
X |
|
|
|
X |
|
54201 |
23279 |
|
|
X |
X |
|
|
|
|
X |
55202 |
23279 |
|
|
X |
|
X |
|
X |
|
|
54210 |
23279 |
|
|
X |
|
X |
|
|
X |
|
54211 |
23279 |
|
|
X |
|
X |
|
|
|
X |
55212 |
23279 |
|
|
X |
|
|
X |
X |
|
|
54220 |
23279 |
|
|
X |
|
|
X |
|
X |
|
54221 |
23279 |
|
|
X |
|
|
X |
|
|
X |
55222 |
23279 |
Decal is optional, add to order as required. 3-way solenoid valve supplied with pumps configured for solenoid valve.
Precision Syringe Pump V6P:
| RESOLUTION |
VALVE FEATURE |
DRIVER & CONTROL FEATURE |
PUMP PT# |
DECAL PT# |
| 48K |
24K |
ROTARY |
DRIVER & CONTROL BOARD |
| X |
|
X |
X |
55023 |
23279 |
|
X |
X |
X |
55123 |
23279 |
Ancillary items:
| DESCRIPTION |
PART NUMBER |
| Communication cable RS485, 6” long; pump to pump (2 units) |
17736 or 17736X |
| Cooling Fan |
24290 |
| Starter Kit (includes all items listed below) |
23427 |
| • 24VDC Power supply |
23429 |
| • Adapter (card edge to AMP connectors) |
23428 |
| • Connector (card edge with solder pins) |
26875 |
| • Software; operator’s manual; and application notes. |
23422 |
| • Communications cable (RS232 / 5’ long; pump to computer) |
17734 |
Software support:
• KloehnControl (configuration and debug program)
• KSerial (command-line communication utility)
• KCom (machine language helper program)
• KProbe (communication trouble shooting utility)
Valves:
| Configuration |
Orifice |
Distribution |
Non-Distribution |
Port Thread |
| Through mtg block |
0.059 |
19218 |
------ |
1/4-28 UNF |
| 2-WAY |
0.059 |
|
------ |
1/4-28 UNF |
| 3-WAY |
0.059 |
17616 |
17615 |
1/4-28 UNF |
| 4-WAY |
0.059 |
17617 |
17712 |
1/4-28 UNF |
| 5-WAY |
0.059 |
17618 |
------ |
1/4-28 UNF |
| 6-WAY |
0.059 |
17619 |
------ |
1/4-28 UNF |
| 8-WAY |
0.059 |
17620 |
------ |
1/4-28 UNF |
| 12-WAY |
0.040 |
24105 |
------ |
1/4-28 UNF |
| 3-WAY |
0.076 |
18189 |
18192 |
1/4-28 UNF |
| 4-WAY |
0.076 |
18190 |
18191 |
1/4-28 UNF |
| 5-WAY |
0.076 |
18188 |
------ |
1/4-28 UNF |
| 6-WAY |
0.076 |
18193 |
------ |
1/4-28 UNF |
| 8-WAY |
0.076 |
17877 |
------ |
1/4-28 UNF |
Valve construction: • V6 pump uses the same valves as the 50300 pump • Valve wetted surfaces have a Kel-F body insert and PTFE (Teflon) plug. • Port connection is 1/4-28 UNF.
• Some valve ports may not be accessible when used with certain syringes sizes. See Application Notes for further information.
Standard Syringe Assembly (individually boxed):
| Size |
Orifice |
Standard Syringe |
ZDV |
Port Thread |
| 25ul |
0.027 |
17591 |
------ |
1/4-28 UNF |
| 50ul |
0.024 |
17592 |
------ |
1/4-28 UNF |
| 100ul |
0.032 |
17593 |
------ |
1/4-28 UNF |
| 250ul |
0.039 |
17594 |
19509 |
1/4-28 UNF |
| 500ul |
0.076 |
17595 |
19538 |
1/4-28 UNF |
| 1.0ml |
0.076 |
17596 |
25429 |
1/4-28 UNF |
| 1.25ml |
0.076 |
17597 |
25431 |
1/4-28 UNF |
| 2.5ml |
0.076 |
17598 |
19539 |
1/4-28 UNF |
| 5.0ml |
0.076 |
17599 |
18463 |
1/4-28 UNF |
| 10.0ml |
0.076 |
17600 |
18469 |
1/4-28 UNF |
| 25.0ml |
0.076 |
17601 |
23734 |
1/4-28 UNF |
| 50.0ml |
0.076 |
17602 |
------ |
1/4-28 UNF |
Syringe construction:
• Syringes have Borosilicate glass. The 25ul, 50ul, and 100ul syringes use a standard plunger tip, while the 250ul through 50ml incorporate a canted spring design.
UHMW Syringes (individually boxed):
| Size |
Orifice |
Standard Syringe |
ZDV |
Port Thread |
| 25ul |
0.027 |
------ |
------ |
1/4-28 UNF |
| 50ul |
0.024 |
24681 |
------ |
1/4-28 UNF |
| 100ul |
0.032 |
24518 |
------ |
1/4-28 UNF |
| 250ul |
0.039 |
19513 |
------ |
1/4-28 UNF |
| 500ul |
0.076 |
24694 |
25427 |
1/4-28 UNF |
| 1.0ml |
0.076 |
24690 |
25413 |
1/4-28 UNF |
| 1.25ml |
0.076 |
------ |
25438 |
1/4-28 UNF |
| 2.5ml |
0.076 |
24685 |
25388 |
1/4-28 UNF |
| 5.0ml |
0.076 |
18857 |
24691 |
1/4-28 UNF |
| 10.0ml |
0.076 |
19110 |
24139 |
1/4-28 UNF |
| 25.0ml |
0.076 |
24688 |
25380 |
1/4-28 UNF |
| 50.0ml |
0.076 |
------ |
------ |
1/4-28 UNF |
Syringe construction:
• Syringes have Borosilicate glass. The 25ul, 50ul, and 100ul syringes use a standard plunger tip, while the 250ul through 50ml incorporate a canted spring design.
• Wetted materials: Borosilicate glass; KEL-F; UHMW


The V6 syringe line is designed to couple to a Syringe V6 Pump and standard valve assembly. These syringes are typically used in various types of instruments and analyzers serving clinical, medical, and analytical instrumentation applications. An array of syringe volumes are available that can be dispensed and/or aspirated over a full 6cm stroke.
For additional custom modifications, see your Kloehn Sales Representative.
Physical
• Length (Dispensed): 4.370”
• Thread Type: 1/4-28 UNF-2A
• Test Pressure: 80 psi
Environmental
• Operating Temperature of 10 to 40°C
• Storage Temperature of -25 to 85°C
• Relative Humidity: Up to 100%
Mechanical
| Syringe Sizes |
10ul |
25ul |
50ul |
100ul |
250ul |
500ul |
1.0ml |
| Orifice Diameter (in) |
.016 |
.016 |
.024 |
.032 |
.039 |
.076 |
.076 |
| Syringe Diameter (in) |
0.4 |
0.4 |
0.4 |
0.4 |
0.4 |
0.4 |
0.4 |
| *Max Drag Force (lbs) |
PTFE |
3.0 |
3.0 |
3.0 |
3.0 |
3.0 |
3.0 |
3.0 |
| UHMW |
7.0 |
7.0 |
7.0 |
7.0 |
7.0 |
7.0 |
7.0 |
| Syringe Sizes |
1.25ml |
2.5ml |
5.0ml |
10ml |
25ml |
50ml |
| Orifice Diameter (in) |
.076 |
.076 |
.076 |
.076 |
.076 |
.076 |
| Syringe Diameter (in) |
0.4 |
0.5 |
0.6 |
0.75 |
1.12 |
1.5 |
| *Max Drag Force (lbs) |
PTFE |
3.0 |
3.0 |
6.0 |
6.0 |
8.0 |
8.0 |
| UHMW |
7.0 |
7.0 |
10.0 |
10.0 |
15.0 |
15.0 |
* Add 1 pound drag force to non-lubricated plunger tip requests.
Chemical
• Standard Syringe Seal: PTFE
• Wetted Materials: Borosilicate Glass, Kel-F, PTFE (or UHMW) (smaller sizes ZDV tips are Teflon coated 316L)
• Syringe tips lubricated with laboratory-grade Silicone
|
|
 |
|
|
|